Download a Monthly QA Report

The Oklahoma Mesonet's quality-assurance (QA) consists of four principal components:
1) laboratory calibration,
2) on-site intercomparison,
3) automated QA, and
4) manual QA.

In the instrument laboratory, all sensors are calibrated to validate or improve upon factory calibrations.

On an annual basis, the accuracy of several types of sensors in the field is verified through comparison to calibrated reference sensors housed in a portable system.

To make a first pass through the 640,000+ observations that are received at the Mesonet each day, automated QA software includes numerous algorithms to evaluate the data received from remote stations.

Finally, the Mesonet’s QA meteorologists employ numerous manual techniques to complement automated QA including analysis of monthly statistics to detect sensor drift or bias (click on a month and year to the left to download a PDF version of a Monthly Report). In addition to detecting problematic sensors, the QA meteorologists trace the true start time of each problem so that appropriate data can be manually flagged as erroneous. The QA meteorologists are responsible for communicating problems to, and coordinating with, appropriate field technicians to ensure proper resolution (see a list of the current outstanding tickets below).

Current Trouble Tickets in the Oklahoma Mesonet
(For questions regarding these tickets, contact qamgr@mesonet.org)

Station Variable Date Reported Problem Description
Ada WS2M 05-Mar-2010 15:55 Monthly QA indicates that WS2M has developed a starting threshold problem
Arnett TB05 11-Feb-2010 11:38 Monthly QA indicates TB05 has a low bias compared to neighbors and TB10. Please bath test both TB05 and TB10.
Beaver TS05 06-Jan-2010 09:57 Monthly QA indicates that TS05 has developed a low bias. Please bath test TS05 and TS10.
Chickasha WS2M 18-Feb-2010 09:29 Sensor has developed a starting threshold problem
Durant TB10 08-Mar-2010 14:21 Monthly QA indicated that TB10 had developed low bias. Please bath test TB05 and TB10.
El Reno DT05 12-Mar-2010 08:12 Fractional water reporting reporting values near 1.6. ST05 and FT05 reporting values that do not appear to be erroneous. ST05 and FT05 are very close in temperature, heater may not be heating
Freedom TB05 09-Mar-2010 13:31 Monthly QA indicates that TB05 has developed a low bias compared to TB10 and neighbors. Please bath test sensors
Grandfield WS2M 08-Feb-2010 12:51 WS2M has starting threshold problem. Problem traces to ice storm.
Jay TB05 24-Feb-2010 09:13 TB05 much warmer than neighbors and like instruments. Tb05 no longer has a smooth curve, suspect that sensor is shallow.
Ketchum Ranch TS30 15-Feb-2010 17:01 TS30 reporting small dips in data. Data is not smooth, variable through out the day.
Pauls Valley TB05 08-Mar-2010 11:18 Monthly QA indicates that TB05 is much warmer than neighbors. TB05 has a large diurnal oscillation, suspect Bare plot is shallow.
Pawnee TB05 01-Mar-2010 12:43 TB05 much warmer than neighbors and like instruments during the daytime. Suspect that sensor is not at correct depth.
Seiling TS10 02-Nov-2009 15:38 Please verify serial numbers for TB05, TB10, TS05, TS10 during next site visit.
Tahlequah TS30 06-Jan-2010 10:22 Monthly QA indicates that TS30 has developed a high bias. Please bath Test sensor.
Tishomingo TB05 24-Feb-2010 09:38 TB05 is much warmer than neighbors and other levels. Suspect that sensor may be shallow.
Waurika TA9M 19-Feb-2010 08:49 TA9M spikes down to large negative values for extended periods of time.
Wilburton TB05 25-Jan-2010 21:54 TB05 and TB10 have a 1.5 C low bias compared to neighbors. Problem traces to lightning strike; please replace.
Wilburton TB10 25-Jan-2010 21:45 TB05 and TB10 have a 1.5 C low bias compared to neighbors. Problem traces to lightning strike; please replace.
Wister TA9M 04-Mar-2010 08:50 TA9M sometimes reports large negative values for several hours over night, then returns to normal. Please check wire connections. Data manually flagged on as-needed basis.